热电偶/RTD探头校准套件

TPCK Series

Thermocouple/RTD Probe Calibration Kit, TPCK-1D | TPCK Series

Thermocouple/RTD Probe Calibration Kit, TPCK-1D

查看下面所有型号

1年保修
  • System Accuracy Better than 1°C
  • Includes ice point TM and hot point™ References, DMM, Voltage Standard, Reference Thermometer, and Accessories
  • Thermocouple Probe Calibrations up to 900°F
  • Reference Probes Included for J, K, T, E Thermocouples

产品描述

TPCK Series offers a complete calibration system for any thermocouple element up to 482°C (900°F). The TPCK kit comes complete with everything needed to calibrate your probes. Included with each kit is a NIST traceable RTD probe, high precision digital RTD thermometer with 0.001°C resolution, 6-digit precision multimeter, hot point™ dry block calibrator, TRCIII ice point™ reference chamber, thermocouple reference probe assemblies, thermocouple millivolt vs. temperature reference tables, thermocouple lead wire and copper jumper jack. Used in calibration or metrology laboratories, the TPCK series is ideally suited to provide a practical solution for your calibration needs.
The hot point™ calibrator allows calibration of thermocouple and RTD probes quickly and easily. An integral RTD sensor assures high measurement accuracy and repeatability. Used in conjunction with the built-in PID controller, exceptional stability of ±0.3°F or better is attained at temperatures up to 482°C (900°F). The stabilizing block features two 1/4 in. diameter wells and one each of 3/16 , 1/8 and 1/16" diameter 6" deep wells. By placing the standard probe (supplied with the TPCK kit) in the well next to the probe under calibration, a temperature comparison is made.

The TRCIII ice point TM calibrator reference point cell relies on an equilibrium of ice and distilled, deionized water at atmospheric pressure to maintain six reference wells at precisely 0°C. The wells extend into a sealed cylindrical chamber containing the distilled, deionized water whose walls are cooled by thermo-electric elements. The reference wells are maintained at 0°C to within ±0.1°C or better and a stability of ±0.04°C for constant ambient. Supplied with the TPCK Series is a set which locates the thermocouple-copper transition reference junction at the end of an 1/8" diameter sheath for easy insertion into the TRCIII ice point® calibration chamber.

Examples of pricing for popular models is referenced in the "To Order" box. Use the "Part Number Builder" below to get pricing for all available models.
For complete product specifications see the Related Links section at the bottom of this page.
订购 所有价格均为人民币税前价格,不包括13%增值税和运费!(指定型号)
网上下单立享5%折扣!

其他客户同时购买: 这个图标可以点击扩大,跳出的清单包括了其他客户在购买此产品时同时购买的其它产品。

产品编号产品描述
数量
产品描述
Quick Ship
  
TPCK-1D
咨询  
货期: 
 
 
Calibration Kit with RS-232 Interface
Quick Ship
Click to expand/minimize a list of what others bought with theTRP-S
TRP-S
¥7,234.86
货期: 2 周
 
Type S thermocouple reference probe
Quick Ship
Click to expand/minimize a list of what others bought with theTRP-R
TRP-R
¥7,234.86
货期: 2 周
 
Type R thermocouple reference probe
所有金额以RMB显示

产品编号生成器

在此处生成您自己的产品编号
选项说明:


(1) Thermocouple Type 选项:
R for Type R
S for Type S
B for Type B
G for Type G
C for Type C
D for Type D
注: 并非所有组合都有效,请查阅规格表上的有效产品编号。
400-820-0559