 8-Channel Benchtop Digital Meter for T/C and RTD Inputs |  | DP9800 Series |  |  
  |  |  
| 
	|  |  |   
    | 
        
			| 8-Channel Benchtop Digital Thermometer for T/C & RTD Inputs 
  |          
           | Click for larger image. |  |  |  |  |  
	
	
|  |  
	| 
	
		|  | 
    	
      	| The DP9800 Series thermometer
can be used in conjunction with a
PC to provide accurate, versatile
8 channel thermocouple and
Pt100 RTD temperature measurements,
scanning and logging of
measured values. It can also be
used as a “stand alone” indicator/
logger and incorporates a digital
backlit LCD display of measured
temperature.
The built-in self-calibration facility
for the DP9800-TC thermocouple
model is a rapid and convenient
method for on-site calibration
and does not require any
additional equipment other
than a special, external link.
Self-calibration of Pt100 ranges is
equally simple and uses plug-in
precision resistors.
The DP9800 is designed to
provide exceptional stability with
high measurement resolution and
represents an ideal crossover
between plant practicality and
laboratory performance at a very
competitive price.The PC software supplied with
the instrument allows control,
measure and calibration functions.
 
 Specifications
 At an ambient temperature of 20°C
 Input/Ranges: Thermocouple to IEC 584
 Type J: -200 to 750°C (-328 to 1382°F)
 Type K: -200 to 1200°C (-328 to 2192°F)
 Type T: -200 to 350°C (-328 to 662°F)
 Type E: -200 to 900°C (-328 to 1652°F)
 Type N: 0 to 1300°C (32 to 2372°F)
 Type R: 0 to 1760°C (32 to 3200°F)
 Type S: 0 to 1760°C (32 to 3200°F)
 Type B: 300 to 1800°C (572 to 3272°F)
 Pt100 to IEC751, 3-Wire: -200 to 850°C (-328 to 1562°F)
 Note: all inputs are non-isolated and thermocouples must be ungrounded style.
 Accuracy: Thermocouples J, K, T, E, and N, better than ±0.1°C ±0.1% of range zero to span ±0.15% of range -100 to - 200°C (J, K, T, and E)
 Thermocouples: R, S, and B, better than ±0.1°C ±0.15% of range
 Linearization: ±0.05°C
 Pt100 Range: Better than ±0.05°C ±0.1% of range
 Zero Drift: ±0.01% of span per °C
 Span Drift: ±0.01% of span per °C
 Display: LCD, backlight
 Display Resolution: Thermocouple ranges 0.1°C, Pt100 range 0.01°C
 User Interface: Front panel key for channel number or auto-scan selection; PC software for all other functions
 Indication: Channel number, measured temperature (°C or °F)
 Reference Junction: Automatic, accurate reference
 Self Calibration: User facility incorporated; the instrument auto-calibrates on every A/D cycle
 Sensor: Open circuit upscale indication, detection and indication (independent alarms should be used for process safety if required) Ambient Operating Temperature: 0 to 50°C (32 to 122°F)
 Storage Temperature: -20 to 70°C (-4 to 158°F)
 Input Terminations: 8 x thermocouple, mini connectors
 Terminal Blocks: 8 x Pt100
 PC Interface: USB
 Power Supply: 6 Vdc (5.5 to 9.0V) via universal mains adaptor (supplied), 120 to 250V, 50/60 Hz
 Logging Interval: 5 second to 1 hour
 |  |  |  |  |  |  |  | | † Price are shown in RMB. 
 Note: Comes complete with power adaptor, USB cable, software and operator’s manual (on CD).
 
 | 
 |  
	|  |  |  |  | 
 |  |  | 
	
	
		| Related Links - 
		  Manual |  |  
		|  |  
		| 
		     
		      | 
		          
		          
		           
		          
		          
		          
		          
		            | Product Manual |  
		            | Download
		                 DP9800-TC and DP9800-RTD - Temperature Monitor |  
		            | Download
		                 DP9800 Series - Communication Manual |  |  |  
		|  |  |  |